There was no patient involvement.Livanova deutschland manufactures the heater cooler system 3t.The incident occurred in york, pennsylvania.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: error message was reproduced and patient pump 1 would not spin.In order to fix the issue, patient pump 1 was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|