• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Nerve Damage (1979); Paralysis (1997); Transient Ischemic Attack (2109)
Event Date 12/15/2021
Event Type  Injury  
Manufacturer Narrative
This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.Of note, multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique product serial/lot numbers.The baseline gender/age characteristics is male/68 years old.The model listed in the report is a representative of the model family, as there is no specific model listed.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: individualized or fixed approach to pulmonary vein isolation utilizing the fourth-generation cryoballoon in patients with paroxysmal atrial fibrillation: the randomized indi-freeze trial.Europace (2022) 24, 921¿927.Doi: 10.1093/europace/euab305 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Literature was reviewed regarding individualized or fixed approach to pulmonary vein isolation with cryoballoon ablation.The authors described one patient who experienced a transient ischemic attack (tia); the patient developed systems of apraxia and sensibility disorders of their face.An intracerebral computerized tomography (ct) did not detect any intracranial bleeding.All symptoms resolved within less than 24 hours.There were other patients who experienced phrenic nerve injury and all of them were asymptomatic and recovered within six months of follow-up.The status of the catheters and sheaths is unknown.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16682685
MDR Text Key312703826
Report Number3002648230-2023-00159
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age68 YR
Patient SexMale
-
-