BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number T5C4484 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the twist clamp of a peritoneal dialysis (pd) transfer set was unable to fully close; further described as ¿twist clamp of the transfer set was difficult to close and impossible to be closed completely¿.This occurred during use of the device for pd therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h6, h10: h10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak testing, clear passage testing and clamp function testing were performed with no issues noted.Torque engagement testing was performed and the twist sleeve closed as expected.The engage and disengage force to open the twist sleeve met specifications.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|