• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Discomfort (2330)
Event Date 03/15/2023
Event Type  Injury  
Event Description
It was reported that the patient presented to the hospital for a follow-up on 15 mar 2023.During the examination, it was noted that there was pain at the implant site of the implantable cardiac monitor (icm).The physician explanted the icm on 15 mar 2023.The patient was in stable condition.
 
Manufacturer Narrative
The device was received in normal working conditions with battery voltage above elective replacement indicator (eri).Analysis performed, indicated normal device functionality.Longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16686879
MDR Text Key312756101
Report Number2017865-2023-16842
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/05/2022
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000079398
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-