• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE Back to Search Results
Model Number EB-1970K
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2023
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Distal case missing.
 
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the distal body chipped.Based on the result, we concluded that it was caused due to the excessive force applied on the distal body.In addition, our technician confirmed that the lg prong top broken, the remote control buttons perforated, the light guide cable buckled, the objective lens cracked, the insertion flexible tube crushed, the suction arm deformed, the lg prong top loose, the angle wire hard to move, the operation channel (primary) resistance, the u/d knob white marking faded/missing, and the u/d lock lever white marking faded/missing; however, these defects are not the main cause and/or irrelevant to the alleged complaint.In terms of distal body chipped, the possibility of dropping into human body could not be denied.Moreover, based on the technical report, ""hr-rpt-0974(distal end)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO BRONCHOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale 07645
8004315800
MDR Report Key16687970
MDR Text Key312796471
Report Number9610877-2023-53025
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEB-1970K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-