MEDOS INTERNATIONAL SARL FMS VUE II FLUID MANAGEMENT AND TISSUE DEBRIDEMENT SYSTEM; DISTENSION UNIT, FLUID, ARTHROSCOPIC
|
Back to Search Results |
|
Model Number 284004 |
Device Problems
Overfill (2404); Improper Flow or Infusion (2954)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported hat during the procedure the it was reported by a healthcare professional in france that during an unknown procedure on (b)(6) 2023, it was observed that the cylinder of the irrigation pump ran dry on the fms vue ii fluid management and tissue debridement system device.Another like device was used to complete the procedure with an unspecified minimal delay.There were no adverse patient consequences reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: seal damaged.Loose pump head roller.Per service reports, this complaint can be confirmed.The seal were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
|
|