Model Number M4735A
Device Problem Inaccurate Synchronization (1609)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  Injury  
Event Description
It was reported to philips that the device did not synchronize while attempting defibrillation.Additional details have been requested.There was no reported patient death or serious injury, however the reported issue/event impacts or potentially impacts therapy, which could cause or contribute to a death or serious injury if the malfunction were to recur.
 
Manufacturer Narrative
Philips received a complaint on the heartstart xl defibrillator indicating that the device failed to sync during daily routine.There was reportedly no patient involvement.The customer determined there was a malfunction with the battery and replaced it to resolve the reported issue.The fse evaluated the device on site and confirmed no issues were found following battery replacement.The device remains at the customer site and no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key16691513
MDR Text Key312800666
Report Number3030677-2023-01447
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K021453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Device Catalogue NumberM4735A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-