Model Number 2110
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported the device dso (downstream occlusion sensor) failed to calibrate.No patient injury was reported.Additional information reported no patient incident, found during testing.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms#617147.The device was returned for investigation.A visual inspection and functional test were performed.Visual inspection did found the smiths tampered seal label was missing.The customer stated problem was not confirmed and could not be duplicated.The downstream occlusion sensor was replaced as a precaution and a full standard pm was performed.The root cause could not be determined.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16699994
MDR Text Key312872166
Report Number3012307300-2023-03726
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586038778
UDI-Public10610586038778
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130394
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0300-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No