• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD; SCREW, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD; SCREW, FIXATION Back to Search Results
Model Number 47-2484-032-50
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-00960, 0001822565-2023-00961.Concomitant medical products: 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head cat: 47248404550, lot: 65378233.Unknown targeting jig.Cmn femoral nail, ccd 125â°, right, 㸠11.5 mm, 21.5 cm, cat: 47249321011, lot: 3143338.Znnâ¿¢, cmn lag screw, 㸠10.5 mm, 115 mm including set screw, cat: 47248511510, lot: 3093322.Report source: foreign: france.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.Product not returned.
 
Event Description
It was reported that the surgeon was implanting a peritrochanteric nail, after speaking with the control team during the operation, it was noted that the screws had been implanted above the nail holes.The nail was removed and replaced with another back up product during the same surgery.The surgeon is concerned that the targeting instrument had contributed to the failure.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.It was determined that the guide caused the screws to misalign and the screws did not cause or contribute to this event.The event will be investigated on medwatch #0001822565-2023-00961.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was determined that the guide caused the screws to misalign and the screws did not cause or contribute to this event.The event will be investigated on medwatch #0001822565-2023-00961.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD
Type of Device
SCREW, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16701625
MDR Text Key312891111
Report Number0001822565-2023-00959
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00889024093997
UDI-Public(01)00889024093997(17)321010(10)65613735
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K082770
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number47-2484-032-50
Device Catalogue Number47248403250
Device Lot Number65613735
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexMale
-
-