• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CANNULA & CATHETER; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CANNULA & CATHETER; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BE-PAS 2115
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2023
Event Type  malfunction  
Manufacturer Narrative
A follow-up medwatch will be submitted when further information becomes available.
 
Event Description
It was reported that the white part (griff) of the introducer came off during on patient use.Customer use a spare hls cannulae, no harm to patient was reported.(b)(4).
 
Manufacturer Narrative
It was reported that the white part (griff) of the introducer came off during on patient use.Customer use a spare hls cannulae, no harm to patient was reported.The production history record (dhr) of the affected be-pas 2115 with lot# 3000211935 was reviewed on 2023-04-06.According to the dhr result, the product be-pas 2115 passed the defined manufacturing and final release specifications.There is no indication on manufacturing issues.The review of the non-conformities has been performed.It does not show any non-conformity in regards to the batch numbers of reported components and failure.Further, the incoming inspection reports of the affected components griff (batch # 3000188930) and introducer (batch # 3000186977 and # 3000175921) were reviewed on 2023-04-06.The griff was checked visually for particles, pressure marks, rills, streaks, sinks, fat, dirt, blur, cords, scratches, burrs, bubbles and also measured for diameter (inner).The introducer was checked visually for ridges, sharp edges, cracks, streaks, dirt, spots, bubbles and also measured for diameter (outer).All tests were passed as per specifications.The reported failure was identified as part of the current risk management file (dms# (b)(4), v01) and the most probable causes are associated to: 1.Manufacturing: inappropriate assembly of components 2.Logistics: mechanical damage of product due to vibration and impact during transport and storage loosening of handle from introducer due to vibration and impact during transport and storage 3.User: loosening of handle from introducer.Due to the fact that the customer changed the hls cannula with a new one, the complaint could be confirmed.The occurrence rate was calculated for the reported failure and product and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Trackwise # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULA & CATHETER
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key16705241
MDR Text Key313027007
Report Number8010762-2023-00170
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2023
Device Model NumberBE-PAS 2115
Device Catalogue Number701047283
Device Lot Number3000211935
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/13/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/15/2021
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-