• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD SOLIS HPCA PUMPS; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD SOLIS HPCA PUMPS; PUMP, INFUSION Back to Search Results
Catalog Number 21-2112-03
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was a system fault 224, delivered 004.No patient injury reported.Correction: affected item from 21-2111-0300-01 to 21-2112-03.
 
Manufacturer Narrative
This remediation mdr was generated under (b)(4), as a result of warning letter cms#617147.No causes or potential causes of the customers reported problem were found during the review of the service and repair records.A product sample was received for evaluation.Visual inspection showed seal was intact.During functional check, the reported problem was duplicated.Performed process testing according to procedure and failed.Also, running three accuracy tests, the pump was found to be over delivering to the manufacturing specifications.Expulsor was out of mechanical specification causing it to over deliver.Pump's expulsor was trimmed in order to bring the delivery into a more nominal range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS HPCA PUMPS
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16707614
MDR Text Key313037906
Report Number3012307300-2023-03833
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982839
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21-2112-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2022
Was the Report Sent to FDA? No
Date Manufacturer Received04/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-