Model Number 6531
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Event Description
It was reported that device contamination occurred.A 12mm x 4.00mm nc quantum apex was selected for use in a percutaneous coronary intervention of the target lesion located in the left main artery.However, upon opening the package, there was no sterile pouch in the box and the balloon hoop and inflation card were just sitting in the box.The device was not used during the procedure since it was considered unsterile.The procedure was completed with another of same device.There were no patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key16714448
MDR Text Key313053401
Report Number2124215-2023-16335
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729783305
UDI-Public08714729783305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6531
Device Catalogue Number6531
Device Lot Number0028740930
Was Device Available for Evaluation? No
Date Manufacturer Received03/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-