Catalog Number C45101796D-NL |
Device Problems
Break (1069); Moisture Damage (1405); Obstruction of Flow (2423); Moisture or Humidity Problem (2986); Output Problem (3005)
|
Patient Problem
Respiratory Insufficiency (4462)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event; serial #: lot number, expiration date, udi number and device manufacture date: device manufacture date is unknown, no information has been provided to date.510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the customer had issues with the anesthesia circuits that are currently in their circulation.Some have had cracks in the corrugated tubing itself, and some had issues with the dime-sized filter in the circuit.The filter gets soaked with moisture and the thick sponge within the filter gets water-logged and prevents medical staff from ventilating the patient.A patient having resection of mandible for cancer and creation of flap that could not be ventilated due to filter issue.
|
|
Manufacturer Narrative
|
Other text: h6 - updated no product was returned.We are unable to confirm the reported complaint.If the product is returned, icu medical will reopen this complaint for further investigation.A review of the device of the device history records could not be completed as no lot number was provided by the customer.
|
|
Search Alerts/Recalls
|