Model Number 4518
Device Problems Failure to Capture (1081); Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2023
Event Type  malfunction  
Event Description
It was reported that this left ventricular lead exhibited high out of range pacing impedance measurements and failure to capture.No changes have been performed.This lead remains in service.No further adverse patient effects were reported.
 
Event Description
It was reported that this left ventricular lead exhibited high out of range pacing impedance measurements and failure to capture.No changes have been performed.This lead remains in service.No further adverse patient effects were reported.Additional information was received detailing the concerns regarding a possible lead fracture.Further evaluation of the lead was discussed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16732712
MDR Text Key313219818
Report Number2124215-2023-17397
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526417672
UDI-Public00802526417672
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/22/2006
Device Model Number4518
Device Catalogue Number4518
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age91 YR
Patient SexMale
-
-