BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the twist clamp of a peritoneal dialysis (pd) transfer set was unable to fully close.This occurred during use of the device for pd therapy.To resolve this issue, the transfer set was replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Initial reporter facility name: (b)(6).Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however a photograph of the sample was provided for evaluation.The photograph was visually inspected which observed the twist clamp not in the locked position.There was evidence of unknown substance around the notch of the main body that could contribute to the reported cause.Therefore, the reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|