FISHER & PAYKEL HEALTHCARE LTD VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290VX
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in china reported that a mr290vx vented autofeed humidification chamber was found cracked and leaking water during patient use.There was no reported patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290vx vented autofeed humidification chamber was not returned to fisher & paykel healthcare for evaluation.Our investigation is thus based on the information provided by the healthcare facility and our knowledge of the product.Results: the healthcare facility reported that a mr290vx vented autofeed humidification chamber from rt380 adult dual heated evaqua2 breathing circuit was cracked and leaking water during patient use.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.Every mr290vx chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290vx chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290vx vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Event Description
A healthcare facility in china reported that a mr290vx vented autofeed humidification chamber was found cracked and leaking water during patient use.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key16740530
MDR Text Key313318444
Report Number9611451-2023-00321
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290VX
Device Catalogue NumberMR290VX
Device Lot Number2102234120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2023
Initial Date FDA Received04/13/2023
Supplement Dates Manufacturer Received05/16/2023
Supplement Dates FDA Received05/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P RT380 ADULT EVAQUA2 BREATHING CIRCUIT; F&P RT380 ADULT EVAQUA2 BREATHING CIRCUIT
-
-