|
Model Number FGS-0635 |
Device Problems
Positioning Failure (1158); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/30/2023 |
Event Type
malfunction
|
Event Description
|
According to the reporter, there was a capsule which failed to attach to the patient¿s esophagus and still attached to the delivery system upon removal.Another capsule was placed on the same procedure and it attached successfully.There was no patient and user harm, no lubricant was used to facilitate the placement of the capsule.Customer also stated that the spring in the handle popped out.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The capsule was not returned, but the delivery system was available for evaluation.Visual inspection noted the plunger had been rotated more than one-eight of a turn.It was reported that the capsule failed to attach to the patient's esophagus and the capsule was damaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not rotate or otherwise force the plunger beyond the white line on the barrel.Rotating or forcing the plunger beyond this may result in possible damage to the delivery device.This may also interfere with the detachment of the capsule from the delivery device, and cause possible injury to the patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|