• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX XL LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX XL LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1158T/86
Device Problem Failure to Disconnect (2541)
Patient Problems Cardiac Perforation (2513); Pericardial Effusion (3271)
Event Date 03/27/2023
Event Type  Injury  
Event Description
Related manufacturer report number: 2017865-2023-17614, related manufacturer report number: 2017865-2023-17616.It was reported that a patient presented with oversensing noise on the atrial lead resulting in inappropriate mode switch and pacing inhibition.During lead revision, the physician noted the atrial lead, left ventricular (lv) lead, and right ventricular (rv) lead were adhered to each other and elected to extract all three leads.During the procedure, the patient developed pericardial effusion that was promptly repaired.The physician implanted two epicardial leads and completed the procedure.The patient condition was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX XL LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16748198
MDR Text Key313387102
Report Number2017865-2023-17615
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2011
Device Model Number1158T/86
Device Lot Number0002130617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNIFY ASSURA ICD
Patient Outcome(s) Life Threatening;
Patient Age54 YR
Patient SexMale
-
-