SMITH & NEPHEW, INC. LEGION HINGE TIBIAL STABILIZING TOOL; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71434465 |
Device Problems
Break (1069); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal reference number: (b)(4).H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the tip broke off.The broken piece was not returned with the device.The device shows significant signs of wear and use.A functional evaluation reveals due to the device having a piece broke off, this would cause the device not to work as intended.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Event Description
|
It was reported that, during a tka surgery, the legion hinge tibial stabilizing tool device would not turn.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.Visual inspection of the returned device reveals the tip broke off.The broken piece was not returned with the device.The device shows significant signs of wear and use.
|
|
Search Alerts/Recalls
|
|
|