• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A Back to Search Results
Model Number SL-2000M2095D
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2023
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by user facility: detailed inquiry description: foreign object found in the arterial line.The line was not primed or used for the patient.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for further evaluation.Multiple photos were provided for evaluation.A visual evaluation was performed on the photographs, and a black particle inside the arterial line was observed.Visual evaluation of the photographs confirmed the reported defect.A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.The manufacturer evaluated their production processes and were unable to identify a process that would create this defect.Cause could not be determined.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem PA 18018 0027
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethelehem PA 18018 0027
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key16755851
MDR Text Key313450298
Report Number2521402-2023-00086
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046955348763
UDI-Public(01)04046955348763
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSL-2000M2095D
Device Catalogue NumberSL-2000M2095D
Device Lot Number20955023
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-