Model Number N/A |
Device Problems
Difficult to Insert (1316); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/28/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a knee procedure, during patella re-surfacing, the reamer blade would not clear the drill guide.The surgeon noted burring on the inner wall of the drill guide and on the reamer blade.The procedure was completed using an additional set of instruments.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-01066.Reporter source: foreign source: australia customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|