U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Material Integrity Problem (2978); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2023
Event Type  malfunction  
Event Description
It was reported that during a vats procedure, in the middle of surgery, the el5ml wouldn't fire well, while surgeon could still squeeze the handle of device.There was a 'crack' at the body of the device when seeing from the above, therefore, he decided to open a new one.
 
Manufacturer Narrative
(b)(4).Date sent: 4/18/2023.D4: batch # x95u5f.Additional information was requested and the following was obtained: "please clarify how device ¿couldn¿t fire well¿? did device not feed clips? did device feed clips sideways? did device not fire clips (jammed)? if other, please specify.Were there any patient consequences? if yes, please describe: el5ml is packed with pre-loaded clips which cannot feed clips.No further updates from customers on the product complaint." investigation summary.The product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon cycling, the instrument was noted to be empty and locked out.In order to evaluate the condition of the internal components the device was disassembled; upon disassemble the tip of the advancer was found broken and 0 clips were found inside clip track.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16762744
MDR Text Key313937141
Report Number3005075853-2023-02680
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX95U5F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2023
Date Manufacturer Received03/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-