It was reported that during a vats procedure, in the middle of surgery, the el5ml wouldn't fire well, while surgeon could still squeeze the handle of device.There was a 'crack' at the body of the device when seeing from the above, therefore, he decided to open a new one.
|
(b)(4).Date sent: 4/18/2023.D4: batch # x95u5f.Additional information was requested and the following was obtained: "please clarify how device ¿couldn¿t fire well¿? did device not feed clips? did device feed clips sideways? did device not fire clips (jammed)? if other, please specify.Were there any patient consequences? if yes, please describe: el5ml is packed with pre-loaded clips which cannot feed clips.No further updates from customers on the product complaint." investigation summary.The product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon cycling, the instrument was noted to be empty and locked out.In order to evaluate the condition of the internal components the device was disassembled; upon disassemble the tip of the advancer was found broken and 0 clips were found inside clip track.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|