ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315070
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Event Description
The customer complained of discrepant inr results with coaguchek xs meter serial number (b)(4).The meter result was 6.6 inr.The patient was tested again within five minutes and the result was 4.4 inr.The patient's therapeutic range was not provided.
 
Manufacturer Narrative
The meter and test strips were requested for investigation.The returned test strips were measured in comparison with the current test strip master lot.For this purpose two human blood samples from marcumar donors and internal reference meters were used.Donor 1 hct: 46%.Donor 2 hct: 46%.Testing results: donor #1: retention meter and master lot strips: 3.5 inr.Retention meter and customer strips: 3.3 inr.Donor #2: retention meter and master lot strips: 2.5 inr.Retention meter and customer strips: 2.3 inr.All inr values were within the specified target ranges, confirming the functionality of the coaguchek measuring system.No error messages occurred.Returned customer material and retention material comply with specification.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation did not identify a product problem.The cause of the event could not be determined.H3 other text : na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16762980
MDR Text Key313721349
Report Number1823260-2023-01308
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04625315070
Device Lot Number66283811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-