Model Number 283512 |
Device Problems
Overheating of Device (1437); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).E3: reporter is a j&j sales representative.H4: the device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by the sales rep that during an unknown procedure on (b)(6) 2023, it was observed that the fms tornado micro handpiece with buttons device got hot and its suction switch was sticky when moving back and forth.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This complaint involves x (x) devices.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary - the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: ¿ functional : heat ¿ visual : stripped/worn/twisted/cross threaded per service reports, this complaint can be confirmed.The repair of the device was however declined, and it is being placed into long term hold.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review - manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|