• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. PROSTHESIS - UNIVERSAL TITANIUM PROSTHESIS; REPLACEMENT, OSSICULAR PROSTHESIS, TOTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. PROSTHESIS - UNIVERSAL TITANIUM PROSTHESIS; REPLACEMENT, OSSICULAR PROSTHESIS, TOTAL Back to Search Results
Model Number 1150050
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Perforation (2001); Implant Pain (4561)
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the prosthesis implant ¿has been uncomfortable, even painful since its installation in about 2014, importantly it did not improve [their] hearing and ended up projecting through [their] eardrum.¿ the device was explanted, at which time the surgeon stated that ¿it was not connected to the stump of [the] stapes.¿ the patient also reported that their ossicles ¿were removed as a result of a cholesteatoma.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSTHESIS - UNIVERSAL TITANIUM PROSTHESIS
Type of Device
REPLACEMENT, OSSICULAR PROSTHESIS, TOTAL
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key16766593
MDR Text Key313558593
Report Number1045254-2023-00308
Device Sequence Number1
Product Code ETA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K002897
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1150050
Device Catalogue Number1150050
Was Device Available for Evaluation? No
Date Manufacturer Received03/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-