MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Break (1069); High impedance (1291); Unstable (1667); Application Program Problem (2880); Battery Problem (2885); Material Deformation (2976)
|
Patient Problems
Twiddlers Syndrome (4563); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/03/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient's implantable neurostimulator (ins) is at eri.Manufacturer representative (rep) met with patient to check the battery level and discuss replacement alternatives prior to replacement surgery scheduled for (b)(6) , 2023.In interrogating the device, an oor error message popped up.When the impedances were measured, all left brain impedances were normal, but all right brain impedance levels were high (4000 ohlm ranges).No environmental/external/patient factors that may have led or contributed to the issue are known.No troubleshooting was performed.Patient will come to operating room in 10 days, and the impedances will be checked again at the time of replacement.Additional information received from the manufacturer¿s representative (rep) reported at the time of the replacement the surgeon noted the extensions had been twisted multiple times (the patient was flipping/twiddling the implanted device) and the right extension appeared to have an area of insulation that was damaged.The right extension was untangled, but when inserted into the new neurostimulator the impedances were the same as pre-operatively, thus the issue wasn¿t resolved.
|
|
Manufacturer Narrative
|
Mhy product id 37085-40 lot# serial# (b)(4), product type extension product id 37085-40 lot# serial# (b)(4) product type extension section d information references the main component of the system.Other relevant device(s) are: product id: 37085-40, serial/lot #: (b)(4), ubd: 05-may-2015, udi#: (b)(4) ; product id: 37085-40, serial/lot #: (b)(4), ubd: 05-may-2015, udi#:(b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep) reported the neurostimulator was replaced, but the extension that appeared to be damaged wasn¿t explanted.When the new neurostimulator showed the same high impedance measurements, it was clear the issue was not related to the original neurostimulator.The explanted neurostimulator was at eri, and the hospital may or may not choose to send it back.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the implantable neurostimulator (ins) (serial # (b)(6)) determined that the ins was functional and no issues were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the healthcare provider (hcp) reported there was positioning difficulty of the device as thus it was removed.
|
|
Manufacturer Narrative
|
H3.Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|