ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN917285
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2023
Event Type  malfunction  
Event Description
It was reported that "pump had purge failure alarms while on patient.Pump switched out quickly with no patient complications." additional information states, "patient therapy was resumed with no harm to patient".
 
Manufacturer Narrative
(b)(4).Other remarks: n/a.Corrected data: n/a.
 
Event Description
It was reported that "pump had purge failure alarms while on patient.Pump switched out quickly with no patient complications." additional information states, "patient therapy was resumed with no harm to patient".
 
Manufacturer Narrative
(b)(4).Returned for investigation was a pcs assembly.A visual inspection of the pcs assembly was performed and no abnormality was noted.All valves were individually tested by powering them on and off using an external power supply, and all valves functioned as intended.The pcs assembly was installed into a known good lab inventory ac3 for functional testing.The pump was powered on, and pumping was initiated.The purge cycle was performed multiple times with no alarms or errors.The leak test was performed and passed both source side and patient side.The pump was then left to run for over 51 minutes (2 drain cycles) and no alarms or errors occurred.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of "purge failure alarm" is not confirmed.The returned pcs assembly passed visual and functional test specifications during complaint investigation.Based on a review of the dhr, the product met specification upon release.No further action required at this time.Other remarks: n/a corrected data: n/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16768584
MDR Text Key313733471
Report Number3010532612-2023-00216
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10801902172065
UDI-Public10801902172065
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN917285
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received05/30/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-