Model Number 324912
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Event Description
It was reported that a chunk of the bd veo¿ insulin syringe with the bd ultra-fine¿ needle plunger was missing during use.The following information was provided by the initial reporter: "1st plunger issue -chunk out of the plastic missing, was not able to push insulin in.2nd the needle has a burr, extra chunk on needle and cant come off.Some metal deformation.Injury- no".
 
Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a chunk of the bd veo¿ insulin syringe with the bd ultra-fine¿ needle plunger was missing during use.The following information was provided by the initial reporter: "1st plunger issue -chunk out of the plastic missing, was not able to push insulin in.2nd the needle has a burr, extra chunk on needle and cant come off.Some metal deformation.Injury- no.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 08-may-2023.H6: investigation summary: the customer returned (1) 1.0ml g 31 ga 6mm syringe, reporting broken barrel.The syringe was visually examined under microscope and observed no damages to the barrel.A functionality test was performed on the returned sample and no issues related to drawing was observed.Based on the sample received and visual examination, embecta was not able to confirm the reported failure.A review of the device history record was completed for batch# 2227496.All inspections and challenges were performed per the applicable operations qc specifications.Based on the sample received, embecta was not able to duplicate or confirm the customer-indicated failure.The root cause cannot be determined, as the reported failure could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VEO¿ INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16771542
MDR Text Key313613821
Report Number1920898-2023-00239
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382903249121
UDI-Public00382903249121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number324912
Device Catalogue Number324912
Device Lot Number2227496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-