• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2224
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented in clinic for a follow-up.Upon review, it was discovered that the pacemaker exhibited incorrect measurements.It was noted that the pacemaker was incorrectly utilizing lifetime at/af burden.No intervention was performed.There were no patient consequences.
 
Manufacturer Narrative
The reported event of at/af burden alerts not triggering was confirmed.Based on the review of the information provided, the root cause of the event was the timer for the at/af detection was turned off but was not turned back on later; hence the at/af burden was not detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT MRI DR RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16773801
MDR Text Key313637069
Report Number2017865-2023-18001
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2016
Device Model NumberPM2224
Device Lot Number4756007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-