• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION SPECTRUM IQ INFUSION PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION SPECTRUM IQ INFUSION PUMP; PUMP, INFUSION Back to Search Results
Catalog Number 3570009
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a spectrum iq pump had no flow during delivery.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: a service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRUM IQ INFUSION PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDINA
711 park ave
medina NY 14103
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16776908
MDR Text Key313656328
Report Number1314492-2023-01379
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3570009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-