This is filed to report a gripper line break, causing a gripper actuation issue.It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with grade of 4.The first clip (xtw) was implanted with no reported issue.An ntw clip was advanced to the mitral valve.It was noted that one of the grippers would not drop.The grippers were cycled a few times and the gripper started working.On the third grasping attempt, the anterior gripper stopped dropping.It was noted that there was difficulty capturing the leaflets.The clip arms were closed, capturing both leaflets.The clip was implanted and stable on both leaflets.Mr was reduced to grade 3.It seems that the gripper line broke, which caused the anterior gripper not to drop.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
In this case, the reported difficult to open or close associated with single gripper actuation issue and gripper line break could not be replicated in a testing environment due to the returned condition of the device (clip was implanted and therefore was not returned).The reported difficult or delayed positioning associated with leaflet capture could not be replicated in a testing environment due to procedural/operational circumstances.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints from the lot.Based on the results of the analysis and the information provided, a cause for the reported single gripper actuation issue resulting in the difficulty capturing the leaflets (difficult or delayed positioning) cannot be determined.The gripper line break appears to be related to the user¿s perception for the cause of the single gripper actuation issue.There is no indication of a product issue with respect to manufacture, design, or labeling.
|