UNITED STATES ENDOSCOPY GROUP, INC. RAPTOR GRASPING DEVICE Back to Search Results
Model Number 711177
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
The user facility reported via medwatch report (mw5115832) that their raptor grasping device worked on the first pass but would not open during the second pass.No report of injury.
 
Manufacturer Narrative
The device subject of this complaint was not returned to steris endoscopy for evaluation.Without the return of the device for evaluation, the root cause could not be determined.The device history record was reviewed and confirmed the device lot was manufactured to specification.There have been no other complaints associated with this lot.Statements in the instructions for use include: "actuate the device by moving the slider on the handle back and forth to confirm that the grasping jaws open and close smoothly.If the unit does not function properly, or there is evidence of damage (e.G.Bends, kinks, misshapen jaws, misaligned jaws, exposed wires) do not use this product and contact your local product specialist.Do not use excessive force on the handle and do not coil the catheter outside of the endoscope.Excessive force or coiling may damage the device or damage the endoscope and may result in accidental injury to the patient or clinician.The following conditions may cause the device to function improperly: advancing the handle to the open position with too much speed or force.Attempting to pass or open the device in an extremely articulated endoscope.Attempting to actuate the device in an extremely coiled position.Actuating the device when the handle is at an acute angle in relation to the sheath.".No further issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPTOR GRASPING DEVICE
Type of Device
GRASPING DEVICE
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer (Section G)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer Contact
coletta cohara
5976 heisley road
mentor, OH 44060
4403586251
MDR Report Key16780459
MDR Text Key313710943
Report Number1528319-2023-00012
Device Sequence Number1
Product Code OCZ
UDI-Device Identifier00724995183653
UDI-Public00724995183653
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number711177
Device Catalogue Number711177
Device Lot Number4442392
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/24/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-