Model Number 10-85-00
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a report that the flow rate knob of the s5 double roller pump was stuck during a maintenance.There was no patient involvement.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 system.The incident occurred in japan.During maintenance and inspection, it was confirmed that the setting knob of pump "b" was stuck.The pump was sent to livanova japan for investigation and the issue was confirmed.The shaft angle encoder, s5 knob and related necessary parts were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See intial report.
 
Manufacturer Narrative
Review of livanova complaints database revealed no similar event since unit installation in 2016.Based on available information, since the event is isolated, it cannot be excluded that the knob was blocked due to a mechanical defect which also led to the inefficiency of the shaft angle encoder.This did not allow the speed regulation of the roller pump involved.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE ROLLER PUMP 85
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key16783509
MDR Text Key314165415
Report Number9611109-2023-00187
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817900405
UDI-Public(01)04033817900405(11)160809
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K210130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-