LIVANOVA DEUTSCHLAND S5 DOUBLE ROLLER PUMP 85; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/27/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland has received a report that the flow rate knob of the s5 double roller pump was stuck during a maintenance.There was no patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 system.The incident occurred in japan.During maintenance and inspection, it was confirmed that the setting knob of pump "b" was stuck.The pump was sent to livanova japan for investigation and the issue was confirmed.The shaft angle encoder, s5 knob and related necessary parts were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See intial report.
|
|
Manufacturer Narrative
|
Review of livanova complaints database revealed no similar event since unit installation in 2016.Based on available information, since the event is isolated, it cannot be excluded that the knob was blocked due to a mechanical defect which also led to the inefficiency of the shaft angle encoder.This did not allow the speed regulation of the roller pump involved.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Search Alerts/Recalls
|
|
|