BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
|
Back to Search Results |
|
Model Number 3851 |
Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a hole in the shaft occurred.A 10mmx2.50mm was selected for use.During the procedure, it was noted that the wire came out from the part other than the wire exit port.The procedure was completed with another of the same device.There were no patient complications reported post procedure.
|
|
Event Description
|
It was reported that the wire came out from the part other than the wire exit port.A 10mmx2.50mm was selected for use.During the procedure, it was noted that the wire came out from the part other than the wire exit port.The procedure was completed with another of the same device.There were no patient complications reported post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was in a deflated state.No issues were noted with the balloon of the device that may have potentially contributed to the complaint incident.The markerbands and blades and tip section of the device were visually and microscopically examined, and no issues were noted with the markerbands, blades or tip of the device that may have potentially contributed to the complaint incident.All blades were present and fully bonded to the balloon material.A visual and tactile examination of the hypotube identified no kinks or damages.A visual and microscopic examination identified bunching to the wire lumen.This bunching was located at 5cm proximal from the tip.A microscopic examination of the lumen identified clear bunching in the inner/wire lumen.Further microscopic examination of the extrusion found a perforation in the outer lumen, at the site of the bunched inner.Due to the lumen damage, it was not possible to pass a recommended 0.014inch size wire through the wire lumen.
|
|
Search Alerts/Recalls
|
|
|