• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. ORBERA 365 INTRAGASTRIC BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. ORBERA 365 INTRAGASTRIC BALLOON Back to Search Results
Model Number B-50012
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
Patient Problems Abdominal Pain (1685); Failure of Implant (1924); Perforation (2001)
Event Date 02/13/2023
Event Type  malfunction  
Manufacturer Narrative
Initial medwatch submitted to the fda on 21/apr/2023.A review of the device labeling notes the following: risk review verbiage less last including labeling and monitoring.The current orbera365¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential event of pain (abdominal); hospitalization, stomach perforation / erosion and early removal as follows: the igb is composed of a soft silicone elastomer and is easily damaged by instruments or sharp objects.The igb must be handled only with gloved hands and with the instruments recommended in this document.Each physician and patient should evaluate the risks associated with endoscopy and intragastric balloons (see complications below), and the possible benefits of a temporary treatment for weight loss prior to use of the orbera365¿ system.To prevent ulcers and control gastroesophageal reflux symptoms, it is recommended that the patient start a program of oral proton pump inhibitors (ppis) for approximately 3-5 days, prior to igb placement so a maximal gastric acid suppression effect will be present on the day of placement.It is recommended that the ppi dose be given sublingually after igb placement if nausea and/or vomiting are present.A starting full dose daily regimen of an oral ppi should be continued as long as the igb is in place.Other medications that are started prophylactically should be continued after igb placement until they are no longer needed.Furthermore, subjects will be directed to avoid medications known to cause or exacerbate gastroduodenal mucosal damage.The risk of intestinal obstruction may be higher in patients who have had prior abdominal or gynecological surgery.The risk of intestinal obstruction may be higher in patients who have a dysmotility disorder or diabetes.The physiological response of the patient to the presence of the orbera365¿ system balloon may vary, depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements, and the overall diet of the patient may also affect the response.Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, acute pancreatitis, spontaneous inflation, ulceration, gastric and esophageal perforation, and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms.Possible complications of the use of the orbera¿ system include: death due to complications related to intestinal obstruction, gastric perforation, is possible.Pregnancy or breast-feeding contraindicates use of this device.Should pregnancy be confirmed at any time during the course of treatment, the device should be removed as soon as it is safely possible.Gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus.Spontaneous over inflation of an indwelling balloon with symptoms including intense abdominal pain, swelling of the abdomen (abdominal distension) with or without discomfort, difficulty breathing, and/or vomiting.Patients experiencing any of these symptoms should be counseled to seek immediate care.Patients with an igb that present with severe abdominal pain that have a negative endoscopy and x-ray may additionally require a ct scan to definitively rule out a perforation.Note that, continued nausea and vomiting could result from direct irritation of the lining of the stomach, as a result of the balloon blocking the outlet of the stomach, or hyperinflation of the balloon.The labeling is adequate as it addresses the reported complaint.The occurrence of this reported complaint for this product will continue to be monitored as appropriate.Additional information: dhr review was completed for lot number af04849.The subject product met all specifications and requirements in effect at the time of manufacture.There are no other complaints against this lot number and allegation.
 
Event Description
Patient experienced diffuse abdominal pain, perforation and fainting.Emergency surgery was performed.Patient is doing well.
 
Manufacturer Narrative
Supplement #01 medwatch submitted to the fda on 23/jun/2023.Additional information: dhr review was completed for lot number, af04849.The subject product met all specifications and requirements in effect at the time of manufacture.There are no other complaints against this lot number and allegation.Device evaluation summary: assessment of the device involved in this complaint was not possible.The user effect deflation is a known and labeled possible event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA 365 INTRAGASTRIC BALLOON
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg. 1 ste. 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
bldg b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
adriana russell
1120 s. capital of texas hwy
bldg. 1 ste. 300
austin, TX 78746
5122795114
MDR Report Key16788524
MDR Text Key313755772
Report Number3006722112-2023-00078
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020725
UDI-Public(01)10811955020725(17)20231220(10)AF04849
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/20/2023
Device Model NumberB-50012
Device Catalogue NumberB-50012
Device Lot NumberAF04849
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient SexFemale
-
-