Model Number C-VH-3010
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2023
Event Type  malfunction  
Manufacturer Narrative
Trackwise id: (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure the t.W.Power supply connector at the bottom gets hot, they swapped out cables to isolate problem to the actual power supply.They pulled it out of use.They used another vh-3010 to complete the procedure.No patient effects.
 
Manufacturer Narrative
Trackwise (b)(4) updated sections: b4, d10, g4, g7, h2, h3, h6, h10 the reported device is an oem device.The certificate of conformance was reviewed for the serial (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text : device not returned.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key16791488
MDR Text Key313812650
Report Number2242352-2023-00314
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2018
Device Model NumberC-VH-3010
Device Catalogue NumberC-VH-3010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN.
-
-