Model Number HARH36
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2023
Event Type  malfunction  
Event Description
It was reported that during an endoscopic kidney surgery, before used on the patient, the package was found damaged.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Manufacturer Narrative
(b)(4).Date sent: 4/24/2023.D4 batch #: x95a6t.Additional information was requested and the following was obtained: does the damage to the package compromise the sterility of the device?--unk.Investigation summary the product was returned for evaluation.Visual inspection was conducted on the returned device.The device was returned sealed inside its original packaging.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; it was noted to have a hole in the tyvek, like a scratch.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch x95a6t, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16795118
MDR Text Key313819069
Report Number3005075853-2023-02792
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014621
UDI-Public10705036014621
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHARH36
Device Catalogue NumberHARH36
Device Lot NumberX95A6T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2023
Date Manufacturer Received04/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-