• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number ALLURA XPER FD20
Device Problems Unstable (1667); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2023
Event Type  malfunction  
Event Description
It has been reported to philips that on (b)(6) 2023, at the start of a procedure, when moving the c-arm a loud bang was heard and the c-arm suspension seemed to become unstable.The procedure was stopped.No patient harm has been reported.A philips engineer evaluated the system onsite and identified two cracks in the flexmove ceiling carriage.The ceiling carriage is divided into two sections.The observed cracks were at the connection between the two sections.Onsite evaluation also identified that out of the 21 screws of the flexmove ceiling carriage rail, 8 were broken and 9 were loose.  the customer is requested to not use the system until the system is repaired.The investigation is ongoing.A follow up report will be submitted when further information is available.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue occurred.The philips field service engineer (fse) inspected the system onsite and confirmed that the system was unable to move laterally using motorized movement and had a loud bang when moving laterally manually.A review of the system log file showed that there was a collision issue.The fse replaced the flex move carriage.After the replacement of the flex move carriage, the system was returned to use in good working order.These codes were updated based on investigation outcome.The evaluation method code grid was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16797702
MDR Text Key313848207
Report Number3003768277-2023-02481
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838054202
UDI-Public00884838054202
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K161563
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722028
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-