Device evaluated by manufacturer: the device was returned for analysis.Visual inspection revealed that only the main coil was returned.The main coil was found stretched, kinked and detached at the interlocking arm.Since only the main coil returned, the functional inspection could not be performed.Microscope inspection revealed the main coil was detached.
|
Reportable based on device analysis completed on 05apr2023.It was reported that the coil prematurely detached.The target lesion was located in the non-calcified internal iliac artery.A 6mm x 20cm interlock-35 was selected for use.During the procedure, it was noted that after placing the four interlocks, an attempt was made to insert this device; however, it stopped advancing from the middle of the catheter, and the coil detached while pushing back.Since the coil was visible through the catheter, only the coil was remove and the procedure was completed with another of the same device.No patient complications reported.However, device investigations revealed that the coil was detached at the interlocking arm.
|