• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number JC4040
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a cysto/bladder irrigation set had a black mark on the inner portion of the tubing which looked burnt.This issue was discovered when opening the device package.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
D4: expiration date: january 2028.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which revealed a brown particle embedded between the walls of the tubing, which was noted to be a degraded piece of burned plastic.Due to the nature of the sample no additional tests could be performed.The reported condition was verified.The cause of the reported condition was a manufacturing related issue.This issue is not likely to cause or contribute to a death or serious injury as particulate matter outside of the fluid path does not impact the product delivered to the patient.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOSCOPY IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16809908
MDR Text Key313959553
Report Number1416980-2023-01987
Device Sequence Number1
Product Code LJH
UDI-Device Identifier00085412042695
UDI-Public(01)00085412042695
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJC4040
Device Lot NumberR23A19062
Was Device Available for Evaluation? No
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-