BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL
|
Back to Search Results |
|
Catalog Number JC4040 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a cysto/bladder irrigation set had a black mark on the inner portion of the tubing which looked burnt.This issue was discovered when opening the device package.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
D4: expiration date: january 2028.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which revealed a brown particle embedded between the walls of the tubing, which was noted to be a degraded piece of burned plastic.Due to the nature of the sample no additional tests could be performed.The reported condition was verified.The cause of the reported condition was a manufacturing related issue.This issue is not likely to cause or contribute to a death or serious injury as particulate matter outside of the fluid path does not impact the product delivered to the patient.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|