Customer reported that during a procedure (approximately couple weeks before (b)(6) 2023), the endoflator 50 unit shut off by itself, and they had to restart the unit to move forward with the procedure.Customer stated that the procedure was completed without any patient impact.However they do not have further details regarding this procedure.
|
Per evaluation findings by the manufacturer: the problem described by the customer could be reproduced.The problem occurs when the patient tube is placed on the instrument (nozzle 1.7mm) which is connected to the abdomen during the insufflation break.In this case, the flow only reaches a value of approx.1-3 l/min, which leads to the pressure being reached slowly or not at all (depending on the leakage).The problem can also occur if the leakage at the patient side is similar to the insufflation from the unit to the patient.The unit operates with a maximum insufflation pressure of 50mmhg, which means that the maximum flow is dependent on the instrument used.Therefore, if the instrument resistance becomes higher, the set flow will no longer be reached until the instrument resistance decreases.If the insufflation pressure isn't reached, the user is advised to replace the instrument, increase the pressure, check the plug connections and replace the tube / filter.Since the user has not performed all of these steps, it could be confirmed that the issue is based on a use error.Further, contamination was found in the high-pressure regulator.The substance of the contamination cannot be identified exactly.It could be detachment from the supply hose.This type of contamination can contaminate the sinter filter and cause the unit to malfunction.However, it seems that the contamination isn't responsible for the problem described by the customer.This issue is also based on a use error.Device was shipped to the customer on march-18-2022 under outbound delivery (b)(4) and had been in use for approx.12 months; device was never returned for service.
|