• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Model Number 8002950
Device Problem Defective Device (2588)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Other, other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Event Description
It was reported that the air in line sensor went out.No patient involvement reported.
 
Manufacturer Narrative
Other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one device was received for investigation.During the visual inspection, the air detector sensor is loose, and can be wiggled back and forth.This can possibly cause it to be more sensitive and alarm more often.The device has an old design clear float switch which is known to crack/leak.The membrane switch is damaged (has some burn marks) but was still operational.During the functional test, it was found that the sensor works as intended but it was loose.The root cause was found to be that the screws securing sensor to front cover of air detector were not properly secured.The sensor screws were tightened.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date of 2012-06-14 and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Manufacturer Narrative
Health effects and health impact updated.
 
Event Description
Additional information received stating that the problem was found while preparing for patient use.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16820276
MDR Text Key314085248
Report Number3012307300-2023-04629
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8002950
Device Catalogue Number8002950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-