• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1456Q/86
Device Problems Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Perforation (2513); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented during implant procedure.During procedure, perforation was noted before the left ventricular lead was implanted.The reported lead was not implanted.The patient was in stable condition.
 
Event Description
Additional information received noted that the lead perforated during implant procedure while the guidewire was inserted, causing insulation damage.The patient did not experience perforation was stable after the procedure.No further interventions were performed.
 
Manufacturer Narrative
Correction: b1, b2, b5, h1 - patient perforation was erroneous reported in the lead was perforated and there was no perforation noted on the patient.The reported event of lead insulation damage was confirmed.As received, a complete lead was returned in one piece for analysis.Visual inspection found a small hole on the lead body insulation consistent with damage caused by puncture of the wire.No other anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16820864
MDR Text Key314073838
Report Number3006705815-2023-02559
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000136914
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-