• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Tissue Injury (4559)
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that an unspecified patient injury occurred in connection with mayfield modified skull clamp (a1059).Additional information has been requested.
 
Manufacturer Narrative
The mayfield skull clamp (a1059) was returned for evaluation: failure analysis : the reported complaint was confirmed from the evaluation.The ratchet extension had a protruding threaded insert.After the internal thread has been visually checked and no defects can be seen on the aluminum frame, the thread insert was replaced, the 80 lb.Jackscrew was checked and the external thread was recut as a precaution.The skull clamp was then subjected to a successful functional check.Root cause : complaint is confirmed via inspection of the unit.The ratchet extension had a protruding thread insert (heli-coil) that required replacement.Additionally, improper or suboptimal positioning of the skull clamp can contribute to patient slippage or movement.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16825828
MDR Text Key314117728
Report Number3004608878-2023-00075
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-