Model Number HL-90 |
Device Problem
Circuit Failure (1089)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event and udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the fluid warmer had a bad power switch.Patient involvement unknown.
|
|
Manufacturer Narrative
|
Device evaluation: one device was received.The visual inspection showed a cracked tank cover and broken power switch.Functional testing was completed and it was confirmed that the device would not power on.The root cause was found to be that the connection between the power switch and pcb is damaged.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date of 2004 and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
|
|
Event Description
|
Additional information received via email.The event was discovered during testing.No medical intervention and no patient injury was reported.
|
|
Manufacturer Narrative
|
Other, other text: b5 and h6.Health effects codes, updated.
|
|