Updated fields: d4, d9, g3, g6, h2, h3, h6, h10 the mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the investigation of the returned skull clamp showed there was play in its pivot lock, it required cleaning, and some worn hardware needed to be replaced for adjustment.To resolve the issues, the swivel lock has been set to a medium torque according to manufacturer specifications, and then the function of the unit was checked for its function.Root cause - the complaint is confirmed via inspection of the unit.Probable root cause is routine use and wear of the unit.Additionally, improper or suboptimal placement of the skull clamp can result in patient movement of slippage.No corrective action has been initiated for the same or similar issue.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|