The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that during stent deployment interaction with the mildly calcified, mildly torturous and 80% stenosed anatomy resulted in preventing the shaft lumens from moving freely; thus resulting in the reported difficult/delayed activation/deployment.Eventually the stent jumped in a spring like release resulting in the reported stent malposition and migration; however, this could not be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
It was reported that the procedure was performed to treat a carotid angioplasty in the internal carotid artery with mild calcification, mild tortuosity and 80% stenosis.A 6-8/40 acculink carotid stent system was deployed; however, it was noted that deployment was delayed.After stent deployment, the stent jumped forward approximately 50% and migration was noted.The procedure was completed with the a new acculink stent.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|