|
Model Number 0087000 |
Device Problem
False Positive Result (1227)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
A review of quarterly events indicated that samples tested on the galileo echo v2.0 automated blood bank system produced inaccurate results for known antibodies in one (1) complaint instance (involving 1 patient sample).A customer reported a rh mistype with anti-d series 4 and anti-d series 5 on echo v2.0 instrument serial number (b)(6).Anti-d4/d5 both showed 1+ reactions but customer believed positive rh is discrepant result and neg rh is expected.Initial sample was resulting as b neg by tube method; weak d in tube was also neg.A second draw from the same patient was also tested on the same echo instrument and by tube method using the same reagents as first used, with both resulting rh negative as expected.Immucor used remote access to review the echo's results from both samples; the initial sample wells appear visually as mixed field.A sample related issue cannot be ruled out (mixed field reactions are a labeled limitation).Per echo limitation: "the galileo echo cannot reliably detect hemagglutination reactions that are graded as 1+ or less in test tube methodology.The galileo echo does not generate an interpretation of mixed-field.Such a mixed-field reaction will be interpreted as positive, negative, or equivocal".
|
|
Manufacturer Narrative
|
No specific root cause or defect was identified; this appears to be related to a mixed-field reaction in the original sample that was resolved by drawing a new sample from the patient.The conclusion(s) code(s) reported herein are assigned according to the facts presented by the affected customer and immucor's assessment and investigation of those facts.When possible, immucor attempts to obtain the actual samples and reagents involved; retention reagents of the same lot that was involved in the event may be used during the investigation if indicated.Also, when possible, immucor uses remote access to review the relevant data archived on the instrument.The events reported on this quarterly malfunction summary report are limited to those in which no patient harm occurred, and no design defect (or other systemic problem) was identified.There is no specific action that users are expected to take to mitigate the reported device failure/malfunction other than to ensure that preventative maintenance is performed as indicated in the instrument instructions.Immucor will continue to track and trend performance and operational issues such as described in this quarterly malfunction summary report.
|
|
Search Alerts/Recalls
|
|
|