|
Model Number 393-800 |
Device Problems
High Readings (2459); Low Readings (2460)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/18/2023 |
Event Type
malfunction
|
Event Description
|
According to the complaint the customer is complaining about deviations on parameter ph, pco2 and po2 measured on the abl800 flex analyzer (serial no.(b)(6)).Values for ph, pco2 and po2 differ between two measurements from the same patient.The customer states that the values measured during the second measurement do not match with the patient's condition.Following table regarding the two measurements carried out: (b)(6).Based on the measurement results the results of 7,33 (ph) and 155 mmhg (po2) measured on the abl800 flex analyzer is considered false high.The 53,2 mmhg (pco2) is considered false low.At a follow up visit radiometer field service engineer cleaned the abl800 flex analyzer and replaced the po2 electrode and pco2 sealing ring and membrane, ph sealing ring and reference membrane.
|
|
Manufacturer Narrative
|
Radiometer has finalized its investigation with the conclusion that the product did not malfunction.No measurement errors were present in either measurement of ph, pco2, po2 or any of the other.Furthermore, all calibrations as well as ac6+ results for ph, pco2 and po2 performed before and after the considered discrepant measurement passed, and nothing extraordinary is observed.
|
|
Search Alerts/Recalls
|
|
|