Visual and dimensional analysis was performed on the returned device.The reported difficulty to insert was unable to be confirmed due to the condition of the returned dc pod.The reported bend/kink and material separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no indication of a lot specific product quality issue.The investigation was unable to determine a conclusive cause for the reported difficulties.It may be possible that the filter was pulled into the pod too quickly or with excessive force causing the noted damage to the delivery catheter pod and preventing the filtration element from being able to load properly; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling.
|