• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Model Number 22437-19
Device Problems Difficult to Insert (1316); Material Separation (1562); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that during preparation of the emboshield nav 6 embolic protection system (eps), the filter could not be inserted into the delivery catheter (dc) pod when using the torque device as it was noted that the tip of the dc was bent and separated in 2 pieces.There was no bunching or wrinkling observed on the dc prior to loading.Another same size emboshield nav 6 was used to complete the procedure.There was no device use or patient involvement.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual and dimensional analysis was performed on the returned device.The reported difficulty to insert was unable to be confirmed due to the condition of the returned dc pod.The reported bend/kink and material separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no indication of a lot specific product quality issue.The investigation was unable to determine a conclusive cause for the reported difficulties.It may be possible that the filter was pulled into the pod too quickly or with excessive force causing the noted damage to the delivery catheter pod and preventing the filtration element from being able to load properly; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16843732
MDR Text Key314280830
Report Number2024168-2023-04488
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137747
UDI-Public08717648137747
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22437-19
Device Catalogue Number22437-19
Device Lot Number2012462
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-